Z-1939-2025 Class II Ongoing

Recalled by Stryker GmbH — Selzach

Recall Details

Product Type
Devices
Report Date
June 18, 2025
Initiation Date
May 22, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product Description

Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S.

Reason for Recall

Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of California and Missouri. The countries of Belgium, Japan, Netherlands.

Code Information

Catalog Number: 1850-0830S. GTIN: 04546540646293. Batch lot number: K09F126