Z-0296-2013 Class II Terminated
FDA device recall Z-0296-2013 was initiated by Zimmer, Inc. on Sep 5, 2012 and is designated Class II. Reason for recall: Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected… The recall status is terminated (terminated Apr 17, 2014). Affected quantity: 204.
Recall Details
- Product Type
- Devices
- Report Date
- Nov 21, 2012
- Initiation Date
- Sep 5, 2012
- Termination Date
- Apr 17, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 204
Product Description
ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.
Reason for Recall
Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank. Based on investigation, this problem appears to be due to a counterfeit chip on the
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Internationally to Belgium, Czech Republic, Egypt, France, Germany, Morocco, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey, and UAE.
Code Information
Catalog 60-3000-101-00, Serial # 3005IACJ, 3006JAEP, 3007CADR, 3007EACT, 3007EAKT, 3007GAGR, 3007HABK, 3007KAAR, 3007KAET, 3008CADE, 3008HAHC, 3008LACK, 3009AACN, 3009HABA, 3009LAEF, 3010CABH, 3010DAHG, 3010GAAC, 3010GAAJ, 3011LABL, 3011LABN, 3011LABQ, 3011LACC, 3011LACE, 3011LACF, 3011LACG, 3011LACH, 3011LACJ, 301 1LACK, 3011LACL, 3011LACM, 3011LACN, 3011LACP, 3011LACQ, 3011LACR, 3011LACS, 3011LACT, 3011LACU, 3011LACW, 3011LADA, 3011LADE, 3011LADF, 3011LADG, 3011LADH, 3011LADJ, 3011LADK, 3011LADM, 3011LADP, 3011LADQ, 3011LADR, 3011LADS, 3011LADT, 3011LADU, 3011LADW, 3011LAEA, 3011LAEB, 3011LAEC, 3011LAED, 3011LAEE, 3011LAEF, 3011LAEH, 3012AAAG, 3012AAAK, 3012AAAM, 3012AAAN, 3012AAAP, 3012AAAQ, 3012AAAR, 3012AAAS, 3012AAAT, 3012AAAU, 3012AAAW, 3012AABA, 3012AABB, 3012AABC, 3012AABE, 3012AABG, 3012AABH, 3012AABK, 3012AABL, 3012AABM, 3012AABN, 3012AABQ, 3012AABR, 3012AABS, 3012AABU, 3012AACA, 3012AACB, 3012AACC, 3012AACD, 3012AACE, 3012AACH, 3012AACJ, 3012AACM, 3012AACN, 3012AACP, 3012AACQ, 3012AACR, 3012AACS, 3012AACT, 3012AACW, 3012AADA, 3012AADB, 3012AADC, 3012AADD, 3012AADE, 3012AADF, 3012AADG, 3012BAAA, 3012BAAB, 3012BAAC, 3012BAAD, 3012BAAE, 3012BAAF, 3012BAAG, 3012BAAH, 3012BAAJ, 3012BAAK, 3012BAAL, 3012BAAM, 3012BAAN, 3012BAAP, 3012BAAQ, 3012BAAR, 3012BAAS, 3012BAAT, 3012BAAU, 3012BAAW, 3012BABA, 3012BABB, 3012BABC, 3012BABD, 3012BABE, 3012BABF, 3012BABG, 3012BABH, 3012BABJ, 3012BABK, 3012BABL, 3012BABM, 3012BABN, 3012BABP, 3012BABQ, 3012BABR, 3012BABS, 3012BABT, 3012BABU, 3012BABW, 3012BACE, 3012BACG, 3012BACH, 3012BACJ, 3012BACK, 3012BACL, 3012BACM, 3012BACN, 3012BACP, 3012BACQ, 3012BACR, 3012BACS, 3012BACT, 3012BACU, 3012BACW, 3012BADA, 3012BADB, 3012BADC, 3012BADD, 3012BADG, 3012BADH, 3012BADJ, 3012BADK, 3012BADL, 3012BADN, 3012BADP, 3012BADR, 3012BADS, 3012BADU, 3012BADW, 3012BAEA, 3012BAEB, 3012BAEC, 3012BAEH, 3012BAEJ, 3012BAEL, 3012BAEM, 3012BAEN, 3012BAEQ, 3012BAER French Catalog 60-3000 301-02 Serial # 3012 AAAE German Catalog 60-300060-301-03, Serial # 3012AAAF Spanish Catalog 60-3000-301-11, Serial # 3012AAAB, 3012AAAC, 3012AACG, 3012BACA, 3012BACB, 3012BACC, 3012BACD Dutch 60-3000-301-12, Serial # 3012AAAA, 3012AAAD, 3012EACA