Z-0328-2024 Class II Ongoing
FDA device recall Z-0328-2024 was initiated by Aizu Olympus Co., Ltd. on Sep 25, 2023 and is designated Class II. Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the… The recall status is ongoing. Affected quantity: 823 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 29, 2023
- Initiation Date
- Sep 25, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 823 units
Product description
Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.
Reason for recall
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Distribution pattern
Worldwide - US Nationwide distribution.
Code information
Model Number (UDI-DI): BF-UC180F (04953170356360), BF-UC190F (04953170399831).