Z-0379-2017 Class II Terminated

Recalled by Abbott Laboratories — Abbott Park, IL

FDA device recall Z-0379-2017 was initiated by Abbott Laboratories on Oct 11, 2016 and is designated Class II. Reason for recall: Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21). The recall status is terminated (terminated May 17, 2017). Affected quantity: 26, 837 kits.

Recall Details

Product Type
Devices
Report Date
Nov 16, 2016
Initiation Date
Oct 11, 2016
Termination Date
May 17, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
26, 837 kits

Product Description

ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.

Reason for Recall

Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).

Distribution Pattern

Nationwide and Worldwide Distribution.

Code Information

Lot: 31497UN14 Exp.: 17 Oct. 2016; Lot: 09353UNI15 Exp.: 12 Feb. 2017; Lot: 45216UN15 12 May 2017; Lot: 14596UN15 Exp.: 30 Jul. 2017; Lot: 37055UN15 Exp.: 12 Oct. 2017; Lot: 14782UN16 Exp.: 21 Apr. 2018