Z-0422-2015 Class II Terminated
FDA device recall Z-0422-2015 was initiated by Hospira Inc. on Mar 19, 2013 and is designated Class II. Reason for recall: Some patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients… The recall status is terminated (terminated Sep 9, 2016). Affected quantity: Not Applicable.
Recall Details
- Product Type
- Devices
- Report Date
- Dec 3, 2014
- Initiation Date
- Mar 19, 2013
- Termination Date
- Sep 9, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- Not Applicable
Product Description
Reason for Recall
Distribution Pattern
Worldwide Distribution - US: Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Croatia, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Icelanofd, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Netherlands, New Zealand, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom