Z-0437-2013 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0437-2013 was initiated by Zimmer, Inc. on Sep 21, 2012 and is designated Class II. Reason for recall: Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating… The recall status is terminated (terminated Apr 17, 2014). Affected quantity: 123 worldwide (33 USA).

Recall Details

Product Type
Devices
Report Date
Dec 5, 2012
Initiation Date
Sep 21, 2012
Termination Date
Apr 17, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
123 worldwide (33 USA)

Product Description

NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee prostheses. The device is indicated for patients with severe knee pain and disability due to rheumatoid arthritis; osteoarthritis; traumatic arthritis; polyarthritis; collagen disorders and/or avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction or prior patellectomy); moderate valgus, varus or flexion deformities; salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.

Reason for Recall

Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset

Distribution Pattern

Worldwide Distribution - USA including AZ,CA, FL, IA, IL, IN, MO, NE, NJ, NY, OH, PA, WA. Internationally to Thailand, Canada, Germany, France, UK, Hungary, Netherlands, Russian Fed, and Sweden,

Code Information

Catalog 00-5986-047-02, lots 62049114 and 62045235.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0437-2013
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.