Z-0447-2024 Class II Ongoing
FDA device recall Z-0447-2024 was initiated by Elekta, Inc. on Oct 25, 2023 and is designated Class II. Reason for recall: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is… The recall status is ongoing. Affected quantity: 47 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 13, 2023
- Initiation Date
- Oct 25, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 47 units
Product description
ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Reason for recall
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Distribution pattern
worldwide
Code information
UDI/DI N/A, Serial Numbers: 152140, 152617, 152691, 152004, 152683, 152994, 153935, 152620, 152809, 152854, 152867, 152885, 152964, 152974, 153192, 153217, 153527, 152499, 153668, 152519, 153118, 153102, 153204, 153166, 152017, 152112, 152325, 152749, 152754, 152755, 152756, 152860, 153006, 151874, 152538, 152839, 152921, 153149, 153267, 152736, 151793, 151802, 151812, 151960, 152600, 152678, 152838
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0447-2024
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026