Z-0472-2025 Class II Ongoing

Recalled by Hobbs Medical, Inc. — Stafford Springs, CT

Recall Details

Product Type
Devices
Report Date
November 27, 2024
Initiation Date
October 4, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product Description

Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4721

Reason for Recall

Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.

Code Information

Catalog Number: 4721; UDI-DI: M84947210; Lot Number: H11-23-057.