Z-0474-2025 Class II Ongoing

Recalled by Hobbs Medical, Inc. — Stafford Springs, CT

FDA device recall Z-0474-2025 was initiated by Hobbs Medical, Inc. on Oct 4, 2024 and is designated Class II. Reason for recall: Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that… The recall status is ongoing. Affected quantity: 12 units.

Recall details

Product Type
Devices
Report Date
Nov 27, 2024
Initiation Date
Oct 4, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product description

Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4718

Reason for recall

Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

Distribution pattern

Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.

Code information

Catalog Number: 4718; UDI-DI: M84947180; Lot Numbers: H08-23-264, H09-23-112.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0474-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0474-2025. Report a data issue: hello@regdatalab.com.