Z-0475-2025 Class II Ongoing

Recalled by Hobbs Medical, Inc. — Stafford Springs, CT

FDA device recall Z-0475-2025 was initiated by Hobbs Medical, Inc. on Oct 4, 2024 and is designated Class II. Reason for recall: Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that… The recall status is ongoing. Affected quantity: 198 units.

Recall details

Product Type
Devices
Report Date
Nov 27, 2024
Initiation Date
Oct 4, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
198 units

Product description

Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4722

Reason for recall

Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

Distribution pattern

Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.

Code information

Catalog Number: 4722; UDI-DI: M84947220; Lot Numbers: H08-23-265, H09-23-113, H10-23-223, H11-23-058.