Z-0482-2013 Class II Terminated

Recalled by Hospira Inc. — Lake Forest, IL

FDA device recall Z-0482-2013 was initiated by Hospira Inc. on Oct 26, 2012 and is designated Class II. Reason for recall: The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may… The recall status is terminated (terminated Nov 14, 2016).

Recall Details

Product Type
Devices
Report Date
Dec 19, 2012
Initiation Date
Oct 26, 2012
Termination Date
Nov 14, 2016
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

Plum A+ Hyperbaric Single Channel Infusion Pumps with Hospira MedNet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature; Hospira, Inc., Lake Forest, IL 60045; pump list number 11005, module list number 11006 Product Usage: The Plum A+ Hyperbaric Infusion System is exclusively designed for hyperbaric therapy in monospace and multiplace chambers.

Reason for Recall

The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise.

Distribution Pattern

Worldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of Australia, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Lavenia, South Africa, Spain, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates and Vietnam.

Code Information

pump list number 11005 and module list number 11006; all serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0482-2013
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.