Z-0492-2020 Class II Ongoing

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-0492-2020 was initiated by Beckman Coulter Inc. on Nov 9, 2018 and is designated Class II. Reason for recall: Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred… The recall status is ongoing. Affected quantity: 83 automate systems.

Recall Details

Product Type
Devices
Report Date
Nov 27, 2019
Initiation Date
Nov 9, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
83 automate systems

Product Description

AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI: 15099590364083 , Model #ODL25255

Reason for Recall

Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .

Distribution Pattern

US: NY, VA OUS: Austria Belgium Czech Republic France Germany Israel Italy Lithuania Mayotte Romania Slovakia South Africa Spain Sweden Switzerland United Kingdom

Code Information

All units containing software Version(s) 4.1.1 and up

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0492-2020
openFDA file date
Oct 21, 2026
Loaded by RegDataLab
Oct 23, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.