Z-0563-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- December 19, 2018
- Initiation Date
- November 2, 2018
- Termination Date
- November 20, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,375 boxes
Product Description
SonoSite L25 Bracket Kit containing 1 Reusable non-sterile bracket, REF numbers P03122-01, P03122-02, P03122-03, P03122-04, and P03122-05, Mfg for: SonoSite, Inc., Bothell, WA.
Reason for Recall
The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.
Distribution Pattern
Foreign distribution was made to Canada, Argentina, Australia, Austria, Belgium, Brazil, Chile, China, Colombia, Croatia, Cyprus, Denmark, Estonia, France, Germany, Hong Kong, India, Israel, Italy, Japan, Latvia, Lebanon, Lithuania, Netherlands, New Zealand, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Trinidad/Tobago, Turkey, Ukraine, United Arab Emirates, and the United Kingdom.
Code Information
Boxes with lot numbers between A000012-A050663, M000880-M995980, and P022503, and all brackets in inventory that no longer have the product labeling.