Z-0585-2015 Class I Terminated

Recalled by Heart Sync, Inc. — Ann Arbor, MI

Recall Details

Product Type
Devices
Report Date
December 31, 2014
Initiation Date
November 11, 2014
Termination Date
November 24, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
sold in boxes of ten electrodes: 1791.3 boxes; 17913 units total

Product Description

ADULT Radiotranslucent Electrode, Part number T100LO-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.

Reason for Recall

Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the

Distribution Pattern

Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.

Code Information

All Lots. Catalog / Part Number T100LO-PHILIPS