Z-0583-2015 Class I Terminated

Recalled by Heart Sync, Inc. — Ann Arbor, MI

FDA device recall Z-0583-2015 was initiated by Heart Sync, Inc. on Nov 11, 2014 and is designated Class I. Reason for recall: Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp… The recall status is terminated (terminated Nov 24, 2015). Affected quantity: sold in boxes of ten electrodes: 4663.8 boxes; 46638 units total.

Recall details

Product Type
Devices
Report Date
Dec 31, 2014
Initiation Date
Nov 11, 2014
Termination Date
Nov 24, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
sold in boxes of ten electrodes: 4663.8 boxes; 46638 units total

Product description

ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.

Reason for recall

Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the

Distribution pattern

Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.

Code information

All Lots. Catalog / Part Number C100-PHILIPS

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0583-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Heart Sync, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Heart Sync, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Heart Sync, Inc. →

Source: openFDA record for Z-0583-2015. Report a data issue: hello@regdatalab.com.