Z-1434-2013 Class II Terminated

Recalled by Heart Sync, Inc — Ann Arbor, MI

Recall Details

Product Type
Devices
Report Date
June 5, 2013
Initiation Date
April 26, 2013
Termination Date
April 16, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17180 units for all part numbers

Product Description

Heart Sync C100-Physio Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible pacing/ cardioversion Defibrillation /ECG Monitoring electrodes Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.

Reason for Recall

On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.

Distribution Pattern

Worldwide Distribution - USA including AL, GA, MI, MN, NC, ND, OH, PA, SD, VA, VT, OH, TX, MA, and Internationally to Korea.

Code Information

Part number : C100-Physio and Lot numbers: Y100112-02, Y102212-03, Y121312-18, Y012413-02, Y022013-01, Y032113-01