Z-0607-2019 Class II

Recalled by GE Healthcare, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
December 19, 2018
Initiation Date
May 24, 2018
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

Discovery NM/CT 670 ES, model 5376204-70-57

Reason for Recall

When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.

Distribution Pattern

CA, DC, OH, UT, WI, Canada

Code Information

GTIN 00840682123877 Serial Numbers: ESDX35013, ESDW35010