Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
December 23, 2020
Initiation Date
November 6, 2020
Termination Date
October 18, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
699 units

Product Description

Tina-quant C-Reactive Protein Gen.3, Catalog number 05172373190 - Product Usage: the in vitro quantitative determination of CRP in human serum and plasma on Roche automated clinical chemistry analyzers.

Reason for Recall

Roche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c701 and 702 modules. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Distribution Pattern

US Nationwide distribution including in the states of AL, CA, FL, GA, HI, IA, IL, IN, KY, MA, MD, MI, MN, MO, NJ, NY, OH, PA, SC, TN, TX, VA.

Code Information

CRPL3 - lot 42402301