Z-0634-2021 Class II Terminated
FDA device recall Z-0634-2021 was initiated by Roche Diagnostics Operations, Inc. on Nov 6, 2020 and is designated Class II. Reason for recall: Roche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c701 and 702 modules. A discoloration (yellow color)… The recall status is terminated (terminated Oct 18, 2024). Affected quantity: 699 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 23, 2020
- Initiation Date
- Nov 6, 2020
- Termination Date
- Oct 18, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 699 units
Product description
Reason for recall
Distribution pattern
US Nationwide distribution including in the states of AL, CA, FL, GA, HI, IA, IL, IN, KY, MA, MD, MI, MN, MO, NJ, NY, OH, PA, SC, TN, TX, VA.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0634-2021
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026