Z-0645-2016 Class II Terminated

Recalled by Cyberonics, Inc — Houston, TX

Recall Details

Product Type
Devices
Report Date
January 27, 2016
Initiation Date
November 18, 2015
Termination Date
September 6, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
29 units

Product Description

Cyberonics Vagus Nerve Stimulator VNS Therapy AspireSR Generator, Model 106

Reason for Recall

Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity.

Distribution Pattern

US Distributions to the states of : IL, NC, KS, and TX., and Internationally to the country of: United Kingdom.

Code Information

Serial numbers 36602, 37019, 39097, 40311, 40465