Z-0659-2016 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- January 27, 2016
- Initiation Date
- December 18, 2015
- Termination Date
- March 2, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,935 units
Product Description
VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.
Reason for Recall
Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.
Distribution Pattern
Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, and United Arab Emirates.
Code Information
All VNS Therapy AspireSR (Model 106) Generators; Device Identifier - (01)05425025750061