Z-0659-2016 Class II Terminated

Recalled by Cyberonics, Inc — Houston, TX

FDA device recall Z-0659-2016 was initiated by Cyberonics, Inc on Dec 18, 2015 and is designated Class II. Reason for recall: Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output… The recall status is terminated (terminated Mar 2, 2017). Affected quantity: 4,935 units.

Recall details

Product Type
Devices
Report Date
Jan 27, 2016
Initiation Date
Dec 18, 2015
Termination Date
Mar 2, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,935 units

Product description

VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.

Reason for recall

Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.

Distribution pattern

Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, and United Arab Emirates.

Code information

All VNS Therapy AspireSR (Model 106) Generators; Device Identifier - (01)05425025750061

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0659-2016
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cyberonics, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0659-2016. Report a data issue: hello@regdatalab.com.