Z-0659-2016 Class II Terminated
FDA device recall Z-0659-2016 was initiated by Cyberonics, Inc on Dec 18, 2015 and is designated Class II. Reason for recall: Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output… The recall status is terminated (terminated Mar 2, 2017). Affected quantity: 4,935 units.
Recall details
- Product Type
- Devices
- Report Date
- Jan 27, 2016
- Initiation Date
- Dec 18, 2015
- Termination Date
- Mar 2, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,935 units
Product description
Reason for recall
Distribution pattern
Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, and United Arab Emirates.