Z-2255-2018 Class II Terminated

Recalled by Cyberonics, Inc — Houston, TX

FDA device recall Z-2255-2018 was initiated by Cyberonics, Inc on Nov 13, 2017 and is designated Class II. Reason for recall: Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead… The recall status is terminated (terminated Oct 31, 2022). Affected quantity: 252.

Recall details

Product Type
Devices
Report Date
Jun 27, 2018
Initiation Date
Nov 13, 2017
Termination Date
Oct 31, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
252

Product description

VNS Therapy Programmer, Model 3000, v1.0 System

Reason for recall

Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).

Distribution pattern

AL, CA, CO, DE, FL, GA, ID, IL, IN, ME, MO, MS, NC, NJ, NY, PA, TN, TN, TX, UT, WA and WI

Code information

Serial Numbers: FMT77F2, FKR57F2, FCR77F2, CJR77F2, BNT77F2, BNS77F2, 7XR77F2, 7XP57F2, 79R77F2, 90S77F2, 97Q57F2, 6NT77F2, 9GP57F2, 6KQ57F2, B7Q57F2, 57P57F2, 55T77F2, J6R57F2, HWR77F2, 1G7G1F2, 1JR57F2, 9XP57F2, 4LR77F2, 4GT77F2, 4GR57F2, 2DT77F2, G2S77F2, FVQ77F2, G5T77F2, FDQ57F2, F5R57F2, F2R57F2, F1Q57F2, GHS77F2, DVR77F2, DNP57F2, DMYB7F2, DJS77F2, D1R57F2, CVQ57F2, H4R57F2, 85S77F2, 89T77F2, 8LYB7F2, 97YB7F2, 6MQ57F2, 6CQ57F2, 5HS77F2, JLQ57F2, JGS77F2, 11R57F2, 1TQ57F2, 1YR77F2, 2DR77F2, ,FXP57F2 , FTR77F2, FPQ57F2, FNQ57F2, FLYB7F2, FFP57F2, GGT77F2, GSQ57F2, CMS77F2, H1R77F2, H3R57F2, 7LS77F2, 9KR57F2, 9NS77F2, 65R57F2, B8R77F2, B9T77F2, 5D0C7F2, B9YB7F2, BJR57F2, JCQ57F2, J9YB7F2, HMYB7F2, CDQ57F2, FXQ57F2, G5S77F2, G3R77F2, FRQ57F2, G7YB7F2, FLS77F2, G90C7F2, FGR77F2, FFT77F2, G9Q57F2, FCT77F2, G9R57F2, F9YB7F2, F9S77F2, F8T77F2, GDQ57F2, F7R57F2, F4T77F2, DWR77F2, DTR77F2, DSQ57F2, GMQ57F2, DLS77F2, DH0C7F2, DGS77F2, DGP57F2, GZR77F2, D1Q57F2, CSQ57F2, CLS77F2, C9YB7F2, C70C7F2, H1R57F2, C1R77F2, 7MS77F2, 78YB7F2, 9SQ57F2, 5VQ57F2, 5RQ57F2, B60C7F2, 5NQ57F2, 5MR77F2, 5KS77F2, B8T77F2, BHP57F2, BJR77F2, 4JS77F2, H4T77F2, JZQ57F2, JMS77F2, JMQ57F2, JG0C7F2, JBQ57F2, J8T77F2, ,HZQ57F2, J7R57F2, HVR77F2, HTP57F2, HNS77F2, HMQ57F2, HKQ57F2, H7S77F2, D3R57F2, 1CQ57F2, 1HT77F2, 1MYB7F2, 2CP57F2, 2LYB7F2, CLYB7F2, CMYB7F2, B4R77F2, B2S77F2, B0R57F2, 9QS77F2, D8Q57F2, 960C7F2, 95T77F2, 8JR77F2, 8HR57F2, BXP57F2, BWR77F2, BTQ77F2, C8YB7F2, C60C7F2, BLR77F2, BLQ57F2, BGR77F2, BGP57F2, CHQ57F2, CKR57F2, CPP57F2, CSR77F2, CTQ77F2, CXR77F2, 9MS77F2, 9JR77F2, 9DQ57F2, D7S77F2, 94T77F2, 8PQ57F2, BWQ77F2, BSR77F2, BNQ57F2, C9R57F2, C8R57F2, C5R57F2, CF0C7F2, C3R57F2, CFP57F2, BDT77F2, B9S77F2, B5R57F2, CPQ57F2, B4R57F2, D2R57F2, 9KQ57F2, D5R77F2, 9F0C7F2, 980C7F2, D9YB7F2, 8WR77F2, DB0C7F2, 9P7G1F2, D3S77F2, D6R57F2, D7R57F2, 9CR77F2, 9G0C7F2, 96R57F2, DFP57F2, C0S77F2, BN7G1F2, BQ7G1F2, C8S77F2, CB0C7F2, CCR77F2, BHR57F2, CHP57F2, BCR77F2, BBR77F2, BB0C7F2, B90C7F2, B70C7F2, B2R57F2, 9VP57F2, 9TQ57F2, 9GR57F2, 9FP57F2, 9BS77F2, ,98R57F2 ,D80C7F2, 970C7F2, D8YB7F2, 95R77F2, 8YR77F2, C1R57F2, C80C7F2, C3R77F2, CNS77F2, 9VQ77F2, D4T77F2, 9DR77F2, 91R57F2, 8VQ77F2, 8RQ57F2, DBS77F2, 8JS77F2, C0R57F2, BTR77F2, BMYB7F2, BKR77F2, BBP57F2 and B8S77F2

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2255-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cyberonics, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2255-2018. Report a data issue: hello@regdatalab.com.