Z-0652-2019 Class II Terminated

Recalled by Ortho-Clinical Diagnostics — Rochester, NY

Recall Details

Product Type
Devices
Report Date
December 26, 2018
Initiation Date
October 19, 2018
Termination Date
January 24, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
682

Product Description

VITROS¿ 3600 Immunodiagnostic System, Version 3.3.1 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid.

Reason for Recall

Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.

Distribution Pattern

US Nationwide Distribution

Code Information

Product Code: 6802783 UDI 10758750002979 Product Code: 6802914 UDI 10758750007103 Serial Numbers: J36000109-J36001169