Z-0657-2019 Class II Terminated

Recalled by Howmedica Osteonics Corp. — Allendale, NJ

Recall Details

Product Type
Devices
Report Date
December 26, 2018
Initiation Date
November 26, 2018
Termination Date
May 26, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
39

Product Description

Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction. The bend at the proximal tip of the ring the discectomy, as well as during the insertion of an IBD during a conventional TLIF procedure.

Reason for Recall

Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe.

Distribution Pattern

Worldwide distribution: US Distribution to the state of: PA and countries of:: Australia, Canada, France, Italy, Netherlands, Poland, and Spain.

Code Information

Lot #s 183432, 183433, 183434, 17E168