Z-0666-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 18, 2024
- Initiation Date
- October 23, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 23
Product Description
MICRO-X Rover Mobile X-ray System, # MXU-RV19
Reason for Recall
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.
Distribution Pattern
US Nationwide Distribution: TX, CA, OH, ID, WA, AZ
Code Information
UDI-DI: 09357123000013. Serial Numbers: 262, 274, 320, 329, 237, 269, 271, 290, 292, 307, 310, 311, 312, 321, 322, 323, 327, 330, 331, 332, 333, 234, 266.