Z-0666-2025 Class II Ongoing

Recalled by Micro-X Ltd. — Tonsley

Recall Details

Product Type
Devices
Report Date
December 18, 2024
Initiation Date
October 23, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23

Product Description

MICRO-X Rover Mobile X-ray System, # MXU-RV19

Reason for Recall

During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.

Distribution Pattern

US Nationwide Distribution: TX, CA, OH, ID, WA, AZ

Code Information

UDI-DI: 09357123000013. Serial Numbers: 262, 274, 320, 329, 237, 269, 271, 290, 292, 307, 310, 311, 312, 321, 322, 323, 327, 330, 331, 332, 333, 234, 266.