Z-1786-2022 Class II Ongoing

Recalled by Micro-X Ltd. — Tonsley, N/A

FDA device recall Z-1786-2022 was initiated by Micro-X Ltd. on Aug 31, 2022 and is designated Class II. Reason for recall: Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure… The recall status is ongoing. Affected quantity: 42 units.

Recall Details

Product Type
Devices
Report Date
Oct 5, 2022
Initiation Date
Aug 31, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 units

Product Description

MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Reason for Recall

Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV, NY, OH, TX and WA and the countries of Australia, France, Israel, Japan, and United Arab Emirates.

Code Information

Model Number: MXU-RV19 UDI Codes: (01)09357123000013(11)210326(21)00222 (01)09357123000013(11)210628(21)00234 (01)09357123000013(11)210629(21)00235 (01)09357123000013(11)210629(21)00236 (01)09357123000013(11)210630(21)00237 (01)09357123000013(11)210708(21)00238 (01)09357123000013(11)210721(21)00240 (01)09357123000013(11)210910(21)00261 (01)09357123000013(11)210913(21)00262 (01)09357123000013(11)210914(21)00263 (01)09357123000013(11)210921(21)00264 (01)09357123000013(11)210922(21)00265 (01)09357123000013(11)210923(21)00266 (01)09357123000013(11)210927(21)00267 (01)09357123000013(11)211006(21)00268 (01)09357123000013(11)211008(21)00269 (01)09357123000013(11)211011(21)00270 (01)09357123000013(11)211011(21)00271 (01)09357123000013(11)211012(21)00272 (01)09357123000013(11)211013(21)00273 (01)09357123000013(11)211013(21)00274 (01)09357123000013(11)211210(21)00288 (01)09357123000013(11)211213(21)00289 (01)09357123000013(11)211213(21)00290 (01)09357123000013(11)220329(21)00305 (01)09357123000013(11)220330(21)00306 (01)09357123000013(11)220331(21)00307 (01)09357123000013(11)220401(21)00308 (01)09357123000013(11)220404(21)00309 (01)09357123000013(11)220405(21)00310 (01)09357123000013(11)220406(21)00311 (01)09357123000013(11)220404(21)00312 Serial Numbers: 222 234 235 236 237 238 240 261 262 263 264 265 266 267 268 269 270 271 272 273 274 288 289 290 305 306 307 308 309 310 311 312

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1786-2022
openFDA file date
Nov 3, 2026
Loaded by RegDataLab
Nov 4, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1786-2022. Report a data issue: hello@regdatalab.com.