Z-0664-2025 Class II Ongoing

Recalled by Micro-X Ltd. — Tonsley

Recall Details

Product Type
Devices
Report Date
December 18, 2024
Initiation Date
May 31, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Reason for Recall

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Distribution Pattern

Worldwide and US Nationwide Distribution: NJ, PR, CA

Code Information

UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿