Z-0682-2025 Class I Ongoing

Recalled by Nova Biomedical Corporation — Waltham, MA

FDA device recall Z-0682-2025 was initiated by Nova Biomedical Corporation on Nov 20, 2024 and is designated Class I. Reason for recall: A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites… The recall status is ongoing. Affected quantity: 347 units.

Recall details

Product Type
Devices
Report Date
Jan 1, 2025
Initiation Date
Nov 20, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
347 units

Product description

StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone meter for near-patient testing

Reason for recall

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

Distribution pattern

Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.

Code information

Model No 63683. UDI-DI (01)38548063685. Software versions: v0.0.13.10 to v0.0.13.44.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0682-2025
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Nova Biomedical Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0682-2025. Report a data issue: hello@regdatalab.com.