Z-0208-2024 Class II Ongoing

Recalled by Nova Biomedical Corporation — Waltham, MA

Recall Details

Product Type
Devices
Report Date
November 8, 2023
Initiation Date
September 18, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
161 units

Product Description

Nova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 63368 Stat Profile Prime Plus POC Analyzer (2) PN 63369 Stat Profile Prime Plus POC Analyzer, Remanufactured P/N: 57400, 63368

Reason for Recall

Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.

Distribution Pattern

Nationwide Foreign: Country Algeria Australia Austria Bangladesh Benin Bolivia Brazil Canada Chile China Colombia Croatia CYPRUS Czech Republic Ecuador Egypt El Salvador Finland France Germany Greece Guatemala India Indonesia Ireland Israel Italy Japan JORDAN Kazakhstan Kenya Kurdistan, Iraq Kuwait Malaysia Malta Mexico Moldova Morocco Nepal Netherlands New Zealand Oman Pakistan Palestine Paraguay Paraquay Peru Philippines Poland Puerto Rico Qatar Republic of Panama Romania Russia Rwanda Saudi Arabia Serbia Slovenia South Africa South Korea Spain Sri Lanka Sudan Sweden Switzerland Syria Thailand Trinidad Trinidad,W.I. Tunisia UK Ukraine United Arab Emerites Uruguay Vietnam

Code Information

UDI-DI: N/A All software versions: International Except China: W1.17.1123.0 and W3.2.1123.0 V1.50.1123.0 China: V1.50.1123.0