Z-0650-2014 Class II Terminated
FDA device recall Z-0650-2014 was initiated by Nova Biomedical Corporation on Oct 10, 2013 and is designated Class II. Reason for recall: Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage The recall status is terminated (terminated Mar 24, 2014). Affected quantity: est. 200,000.
Recall details
- Product Type
- Devices
- Report Date
- Jan 15, 2014
- Initiation Date
- Oct 10, 2013
- Termination Date
- Mar 24, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- est. 200,000
Product description
Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.
Reason for recall
Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage
Distribution pattern
Nationwide Distribution.
Code information
All serial numbers
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0650-2014
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for Nova Biomedical Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.