Z-0735-2023 Class II Ongoing

Recalled by Natus Medical Incorporated — San Diego, CA

Recall Details

Product Type
Devices
Report Date
December 28, 2022
Initiation Date
November 18, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
249 kits

Product Description

The Natus Cranial Access Kit is designed to be a complete, single use and disposable accessory and instrumentation set that is used to create an intracranial burr-hole typically required for procedures such as ventriculostomy or intracranial pressure (ICP) monitoring. The various components of the kit are comprised to make it convenient for the practitioner to have all that they need in one place for fast and efficient intracranial access. cerebral spinal fluid (CSF).

Reason for Recall

Due to Cranial Access Kits containing recalled sterile surgical drapes. The sterile surgical drapes are being recalled due to the liner on the adhesive component is difficult to remove without damaging the product and may render the product unusable.

Distribution Pattern

U.S. Nationwide distribution in the states of AK, CA, CT, DC, FL, GA, IA, ID, IL, KS , MD, MI, MT, ND, NE, NJ, NV, NY, OH, OR, TX, and WA.

Code Information

Part Number: HITH1 UDI-DI Code: 382830048590 Lot Number: 22GDB675 Part Number: HITHNDRZR UDI-DI Code: 382830048620 Lot Number: 22HDA557 Part Number: HITHLT UDI-DI Code: 382830048637 Lot Number: 22IDB724