Z-1363-2016 Class II Terminated

Recalled by Natus Medical Incorporated — Seattle, WA

FDA device recall Z-1363-2016 was initiated by Natus Medical Incorporated on January 8, 2016 and is designated Class II. Reason for recall: NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detachin… The recall status is terminated (terminated February 10, 2017). Affected quantity: 75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units..

Recall Details

Product Type
Devices
Report Date
April 20, 2016
Initiation Date
January 8, 2016
Termination Date
February 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.

Product Description

Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.

Reason for Recall

NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.

Distribution Pattern

Distribution US nationwide and Pakistan.

Code Information

Serial numbers: 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5294; 5293; 5292; 5291; 5290; 5289; 5287; 5283; 5282; 5281; 5280; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; 5129; 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; 5045. Serial numbers were updated in the final status report. They are as follows: 5552; 5551; 5550; 5549; 5510; 5509; 5508;5507; 5506; 5505; 5504; 5389; 5388; 5385; 5383; 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5322; 5294; 5293; 5292; 5291; 5290; 5289; 5288; 5287; 5286; 5285; 5284; 5283; 5282; 5281; 5280; 5279; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; and 5129. The following serial numbers were removed: 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; and 5045.