Z-1363-2016 Class II Terminated

Recalled by Natus Medical Incorporated — Seattle, WA

FDA device recall Z-1363-2016 was initiated by Natus Medical Incorporated on Jan 8, 2016 and is designated Class II. Reason for recall: NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port… The recall status is terminated (terminated Feb 10, 2017). Affected quantity: 75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.

Recall details

Product Type
Devices
Report Date
Apr 20, 2016
Initiation Date
Jan 8, 2016
Termination Date
Feb 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.

Product description

Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.

Reason for recall

NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.

Distribution pattern

Distribution US nationwide and Pakistan.

Code information

Serial numbers: 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5294; 5293; 5292; 5291; 5290; 5289; 5287; 5283; 5282; 5281; 5280; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; 5129; 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; 5045. Serial numbers were updated in the final status report. They are as follows: 5552; 5551; 5550; 5549; 5510; 5509; 5508;5507; 5506; 5505; 5504; 5389; 5388; 5385; 5383; 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5322; 5294; 5293; 5292; 5291; 5290; 5289; 5288; 5287; 5286; 5285; 5284; 5283; 5282; 5281; 5280; 5279; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; and 5129. The following serial numbers were removed: 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; and 5045.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1363-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Natus Medical Incorporated? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Natus Medical Incorporated: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Natus Medical Incorporated →

Source: openFDA record for Z-1363-2016. Report a data issue: hello@regdatalab.com.