Z-1363-2016 Class II Terminated
Recalled by Natus Medical Incorporated
— Seattle, WA
FDA device recall Z-1363-2016 was initiated by Natus Medical Incorporated on Jan 8, 2016 and is designated Class II. Reason for recall: NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port… The recall status is terminated (terminated Feb 10, 2017). Affected quantity: 75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.
Recall details
- Product Type
- Devices
- Report Date
- Apr 20, 2016
- Initiation Date
- Jan 8, 2016
- Termination Date
- Feb 10, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.
Product description
Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.
Reason for recall
NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.
Distribution pattern
Distribution US nationwide and Pakistan.
Code information
Serial numbers: 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5294; 5293; 5292; 5291; 5290; 5289; 5287; 5283; 5282; 5281; 5280; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; 5129; 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; 5045. Serial numbers were updated in the final status report. They are as follows: 5552; 5551; 5550; 5549; 5510; 5509; 5508;5507; 5506; 5505; 5504; 5389; 5388; 5385; 5383; 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5322; 5294; 5293; 5292; 5291; 5290; 5289; 5288; 5287; 5286; 5285; 5284; 5283; 5282; 5281; 5280; 5279; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; and 5129. The following serial numbers were removed: 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; and 5045.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1363-2016
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.