Z-1878-2016 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 8, 2016
- Initiation Date
- April 22, 2016
- Termination Date
- December 16, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 21 filters
Product Description
Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and 100105-01.
Reason for Recall
The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating the interior chamber of the Sterile Drier and its contents.
Distribution Pattern
distributed in the US including AZ, CA, FL, IL, KS, MA, MD, MN, NC, NH, NY, OH, OK, PA, VA, WA, WA DC,and WI and in Canada and Singapore.
Code Information
Filter Serial Numbers: Filter SN: 2522893, 2522897, 2522898, 2522899, 2522900, 2522903, 2522914, 2522915, 2522916, 2522917, 2522918, 2522919, 2522920, 2522921, 2522922, 2522923, 2522924, 2522925, 2522926, 2522930, and 2522931.