Z-1878-2016 Class II Terminated

Recalled by Natus Medical Incorporated — Seattle, WA

FDA device recall Z-1878-2016 was initiated by Natus Medical Incorporated on Apr 22, 2016 and is designated Class II. Reason for recall: The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating… The recall status is terminated (terminated Dec 16, 2016). Affected quantity: 21 filters.

Recall details

Product Type
Devices
Report Date
Jun 8, 2016
Initiation Date
Apr 22, 2016
Termination Date
Dec 16, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21 filters

Product description

Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and 100105-01.

Reason for recall

The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating the interior chamber of the Sterile Drier and its contents.

Distribution pattern

distributed in the US including AZ, CA, FL, IL, KS, MA, MD, MN, NC, NH, NY, OH, OK, PA, VA, WA, WA DC,and WI and in Canada and Singapore.

Code information

Filter Serial Numbers: Filter SN: 2522893, 2522897, 2522898, 2522899, 2522900, 2522903, 2522914, 2522915, 2522916, 2522917, 2522918, 2522919, 2522920, 2522921, 2522922, 2522923, 2522924, 2522925, 2522926, 2522930, and 2522931.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1878-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Natus Medical Incorporated? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1878-2016. Report a data issue: hello@regdatalab.com.