Z-1878-2016 Class II Terminated
FDA device recall Z-1878-2016 was initiated by Natus Medical Incorporated on Apr 22, 2016 and is designated Class II. Reason for recall: The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating… The recall status is terminated (terminated Dec 16, 2016). Affected quantity: 21 filters.
Recall details
- Product Type
- Devices
- Report Date
- Jun 8, 2016
- Initiation Date
- Apr 22, 2016
- Termination Date
- Dec 16, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 21 filters
Product description
Reason for recall
Distribution pattern
distributed in the US including AZ, CA, FL, IL, KS, MA, MD, MN, NC, NH, NY, OH, OK, PA, VA, WA, WA DC,and WI and in Canada and Singapore.