Z-0755-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 6, 2021
- Initiation Date
- December 8, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 units
Product Description
Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001
Reason for Recall
Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
Distribution Pattern
US Nationwide distribution.
Code Information
Serial Numbers: 1925-10; 1931-04; 1931-05;1931-07;1937-10; 1938-01; 1942-10