Z-0771-2013 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0771-2013 was initiated by Zimmer, Inc. on Sep 14, 2012 and is designated Class II. Reason for recall: Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded… The recall status is terminated (terminated Jun 20, 2014). Affected quantity: 275,952 units.

Recall Details

Product Type
Devices
Report Date
Feb 13, 2013
Initiation Date
Sep 14, 2012
Termination Date
Jun 20, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
275,952 units

Product Description

Trilogy F/M Acetabular Shell with Uni-Hole. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.

Reason for Recall

Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.

Code Information

61282267 61633887 61260645 61237937 61500763 61397941 61584769 61199849 61429445 61412089 00369745

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0771-2013
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.