Z-0811-2016 Class II Terminated
FDA device recall Z-0811-2016 was initiated by Verathon, Inc. on Apr 13, 2015 and is designated Class II. Reason for recall: Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such… The recall status is terminated (terminated Oct 26, 2016). Affected quantity: total 9088 units ( 5786 units in the US and 3302 units International).
Recall details
- Product Type
- Devices
- Report Date
- Feb 24, 2016
- Initiation Date
- Apr 13, 2015
- Termination Date
- Oct 26, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- total 9088 units ( 5786 units in the US and 3302 units International)
Product description
GlideScope Video Laryngoscope Reusable Blades (GVL 3, 4, 5 and AVL 2, 3, 4, 5). GVL: The GlideScope Video Laryngoscope is a video laryngoscope that incorporates a high-resolution color camera, an LED light source, a rechargeable lithium battery, and video output to a remote monitor. It consists of a video laryngoscope and a portable video monitor. AVL: The GlideScope AVL Video Monitor system has the same intended use as the GVL System. The difference between systems is that the AVL has a digital platform and the GVL an analog platform.
Reason for recall
Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such that exceeding this number of cycles in the specific chemical agent may affect the potential life of the product.
Distribution pattern
Worldwide Distribution.
Code information
Affected GVL Reusable Blades Serial Numbers, by Part Number GVL 3- 0574-0007: MD131859- MD151622; GVL 4- 0574-0001: LG131994- LG151773; GVL 5- 0574-0030: XL131577- XL151524. Affected AVL Reusable Blades Serial Numbers, by Part Numbers: AVL 2- 0574-0118: AC131852- AC151510; AVL 3- 0574-0115: AD131589- AD151506; AVL 4- 0574-0116: AE131586- AE151558; AVL 5- 0574-0117: AF131503- AF151500.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0811-2016
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab