Z-0811-2016 Class II Terminated

Recalled by Verathon, Inc. — Bothell, WA

FDA device recall Z-0811-2016 was initiated by Verathon, Inc. on Apr 13, 2015 and is designated Class II. Reason for recall: Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such… The recall status is terminated (terminated Oct 26, 2016). Affected quantity: total 9088 units ( 5786 units in the US and 3302 units International).

Recall details

Product Type
Devices
Report Date
Feb 24, 2016
Initiation Date
Apr 13, 2015
Termination Date
Oct 26, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
total 9088 units ( 5786 units in the US and 3302 units International)

Product description

GlideScope Video Laryngoscope Reusable Blades (GVL 3, 4, 5 and AVL 2, 3, 4, 5). GVL: The GlideScope Video Laryngoscope is a video laryngoscope that incorporates a high-resolution color camera, an LED light source, a rechargeable lithium battery, and video output to a remote monitor. It consists of a video laryngoscope and a portable video monitor. AVL: The GlideScope AVL Video Monitor system has the same intended use as the GVL System. The difference between systems is that the AVL has a digital platform and the GVL an analog platform.

Reason for recall

Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such that exceeding this number of cycles in the specific chemical agent may affect the potential life of the product.

Distribution pattern

Worldwide Distribution.

Code information

Affected GVL Reusable Blades Serial Numbers, by Part Number GVL 3- 0574-0007: MD131859- MD151622; GVL 4- 0574-0001: LG131994- LG151773; GVL 5- 0574-0030: XL131577- XL151524. Affected AVL Reusable Blades Serial Numbers, by Part Numbers: AVL 2- 0574-0118: AC131852- AC151510; AVL 3- 0574-0115: AD131589- AD151506; AVL 4- 0574-0116: AE131586- AE151558; AVL 5- 0574-0117: AF131503- AF151500.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0811-2016
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0811-2016. Report a data issue: hello@regdatalab.com.