Z-0821-2019 Class II Ongoing
FDA device recall Z-0821-2019 was initiated by Elekta, Inc. on Dec 13, 2018 and is designated Class II. Reason for recall: There is a potential that Wedge IDs were not included in the DICOM RT PLAN sent from the TPS, which will cause the field in MOSAIQ to be created with no wedge. Treatment of a field without the… The recall status is ongoing. Affected quantity: 3,950 units total (1,747 US).
Recall Details
- Product Type
- Devices
- Report Date
- Feb 20, 2019
- Initiation Date
- Dec 13, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,950 units total (1,747 US)
Product Description
MOSAIQ¿ Oncology Information System and Sequencer MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.
Reason for Recall
There is a potential that Wedge IDs were not included in the DICOM RT PLAN sent from the TPS, which will cause the field in MOSAIQ to be created with no wedge. Treatment of a field without the planned wedge will receive more dose than the treatment plan indicates.
Distribution Pattern
U.S. Nationwide distribution to the following states AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the District of Columbia and Puerto Rico.
Code Information
MOSAIQ Radiation Oncology version 1.0 and higher. Multi-ACCESS version 6.1 and higher. UDI 858164002084 858164002091 858164002107 858164002220 858164002237