Z-0825-2018 Class II Terminated

Recalled by Tosoh Bioscience Inc — Grove City, OH

FDA device recall Z-0825-2018 was initiated by Tosoh Bioscience Inc on Feb 13, 2018 and is designated Class II. Reason for recall: Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone)… The recall status is terminated (terminated Oct 29, 2020). Affected quantity: 5579.

Recall Details

Product Type
Devices
Report Date
Mar 7, 2018
Initiation Date
Feb 13, 2018
Termination Date
Oct 29, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5579

Product Description

ST AIA-PACK¿ PROG II, Progesterone II Assay, Part Number: 025239

Reason for Recall

Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone) supplement.

Distribution Pattern

Distributed in 39 states: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI.