Z-0829-2018 Class II Terminated

Recalled by Fresenius Medical Care Renal Therapies Group, LLC — Waltham, MA

Recall Details

Product Type
Devices
Report Date
March 7, 2018
Initiation Date
October 5, 2017
Termination Date
July 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 machines

Product Description

2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.

Reason for Recall

While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.

Distribution Pattern

US Nationwide Distribution

Code Information

2008T machines: 190766 2008T Hemodialysis System w/Bibag 190895 2008T GEN 2 Bibag without CDX 190766 = 00840861100910 (+Serial Number) 190895 = 00840861100927 (+Serial Number)ux=&Su