Z-3195-2018 Class II Terminated

Recalled by Fresenius Medical Care Renal Therapies Group, LLC — Waltham, MA

FDA device recall Z-3195-2018 was initiated by Fresenius Medical Care Renal Therapies Group, LLC on Jul 13, 2018 and is designated Class II. Reason for recall: In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and… The recall status is terminated (terminated Sep 17, 2019). Affected quantity: 40.

Recall details

Product Type
Devices
Report Date
Sep 26, 2018
Initiation Date
Jul 13, 2018
Termination Date
Sep 17, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40

Product description

2008T HEMODIALYSIS SYSTEM W/BIBAG, Part Number 190766, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.

Reason for recall

In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.

Distribution pattern

US Nationwide in the states of : CA, CT, MI, NC, and NY.

Code information

6T0S160600, 6T0S160624, 6T0S160672, 6T0S160695, 6T0S160732, 6T0S160740, 6T0S160773, 6T0S160781, 6T0S160785, 6T0S160841, 4T0S135962 4T0S135963 4T0S135970 4T0S135972 4T0S135980 4T0S135985 4T0S135987 4T0S136000 4T0S136003 4T0S136070 4T0S136071 4T0S136072 4T0S136077 4T0S136078 4T0S136080 4T0S136081 4T0S136082 4T0S136084 4T0S136088 4T0S136090 4T0S136091 4T0S136106 4T0S136111 4T0S136112 4T0S136114 4T0S136115 4T0S136117 4T0S136120 4T0S136121 4T0S136123

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3195-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Fresenius Medical Care Renal Therapies Group, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-3195-2018. Report a data issue: hello@regdatalab.com.