Z-2151-2019 Class II Terminated

Recalled by Fresenius Medical Care Renal Therapies Group, LLC — Waltham, MA

FDA device recall Z-2151-2019 was initiated by Fresenius Medical Care Renal Therapies Group, LLC on May 29, 2019 and is designated Class II. Reason for recall: Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a… The recall status is terminated (terminated Nov 19, 2020). Affected quantity: 33732973 Cassettes; 18,674 Liberty Cyclers.

Recall details

Product Type
Devices
Report Date
Aug 14, 2019
Initiation Date
May 29, 2019
Termination Date
Nov 19, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33732973 Cassettes; 18,674 Liberty Cyclers

Product description

Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjunction with the Liberty cycler as follows: (1) LIBERTY Line Cycler Set with Stay-safe connectors (Dual Patient Connector) Part Number: 050-87212 (2) LIBERTY Line Cycler Set with Stay-safe connectors (Single Patient Connector) Part Number: 050-87215 (3) LIBERTY Integrated Cycler Set with single Stay-safe patient connector and extended drain line Part Number: 050-87216

Reason for recall

Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination.

Distribution pattern

US Nationwide Distribution

Code information

All lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2151-2019
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Fresenius Medical Care Renal Therapies Group, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2151-2019. Report a data issue: hello@regdatalab.com.