Z-0834-2013 Class II Terminated

Recalled by Ventana Medical Systems Inc — Oro Valley, AZ

FDA device recall Z-0834-2013 was initiated by Ventana Medical Systems Inc on Sep 28, 2012 and is designated Class II. Reason for recall: Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The… The recall status is terminated (terminated Feb 24, 2016). Affected quantity: 54 instruments.

Recall Details

Product Type
Devices
Report Date
Aug 14, 2013
Initiation Date
Sep 28, 2012
Termination Date
Feb 24, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
54 instruments

Product Description

VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.

Reason for Recall

Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu

Distribution Pattern

Worldwide distribution: USA (nationwide) and country of: Germany.

Code Information

Cat#750-601; PN 05987750001; Serial #'s: 310646, 311573, 310477, 311725, 311042, 311623, 312032, 312159, 310615, 310837, 311862, 311043 311572, 311766, 310526, 311011, 311185, 311723, 312082, 310635, 311574, 312084, 310636, 310835, 311179, 310493, 311183, 311769, 311930, 311962, 311963, 310437, 311965, 310997, 311726, 311620, 311512, 311526, 312083, 310468, 310759 (All serial numbers in the market).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0834-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.