Z-0847-2015 Class II Terminated

Recalled by Natus Medical Incorporated — Seattle, WA

Recall Details

Product Type
Devices
Report Date
December 31, 2014
Initiation Date
November 19, 2014
Termination Date
July 20, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1907 total (1413 in US and 494 international)

Product Description

Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.

Reason for Recall

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Distribution Pattern

worldwide

Code Information

All distributed Natus neoBLUE 2 systems with LED PCB replacements received after February 16, 2012. For use only with neoBLUE devices manufactured between 2002 and 2007 (neoBLUE2)