Z-0847-2015 Class II Terminated

Recalled by Natus Medical Incorporated — Seattle, WA

FDA device recall Z-0847-2015 was initiated by Natus Medical Incorporated on Nov 19, 2014 and is designated Class II. Reason for recall: Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity… The recall status is terminated (terminated Jul 20, 2015). Affected quantity: 1907 total (1413 in US and 494 international).

Recall details

Product Type
Devices
Report Date
Dec 31, 2014
Initiation Date
Nov 19, 2014
Termination Date
Jul 20, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1907 total (1413 in US and 494 international)

Product description

Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.

Reason for recall

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Distribution pattern

worldwide

Code information

All distributed Natus neoBLUE 2 systems with LED PCB replacements received after February 16, 2012. For use only with neoBLUE devices manufactured between 2002 and 2007 (neoBLUE2)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0847-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Natus Medical Incorporated? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Natus Medical Incorporated: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Natus Medical Incorporated →

Source: openFDA record for Z-0847-2015. Report a data issue: hello@regdatalab.com.