Z-0868-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 5, 2020
- Initiation Date
- November 16, 2018
- Termination Date
- December 12, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 106 devices
Product Description
Cyberonics VNS Therapy AspireHC Model 105 Generator, 26-0007-4000/3. The firm name on the label is CYBERONICS, INC., Houston, TX. Vagus nerve stimulator.
Reason for Recall
This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.
Distribution Pattern
Distribution US nationwide, including Puerto Rico. There was no foreign/military/government consignees.
Code Information
Serial numbers 35794, 37762, 37792, 37872, 37874, 37881, 37885, 37889, 37890, 37893, 37913, 37915, 37916, 37965, 37975, 37977, 37984, 37987, 37988, 37990, 38051, 38054, 38068, 38132, 38148, 38153, 38154, 38155, 38162, 38173, 38210, 38224, 38226, 38239, 38240, 38243, 38245, 38246, 38248, 38249, 38252, 38268, 38272, 38273, 38274, 38275, 38277, 38279, 38282, 38284, 38292, 38293, 38296, 38299, 38302, 38306, 38308, 38312, 38314, 38315, 38319, 38321, 38326, 38330, 38344, 38443, 38464, 38490, 38505, 38517, 38529, 38530, 38549, 38561, 38564, 38565, 38625, 38628, 38651, 38661, 38689, 38691, 38698, 38702, 38719, 38722, 38730, 38735, 38745, 38751, 38754, 38755, 38771, 38773, 38775, 38779, 38800, 38804, 38805, 38809, 39334, 39344, 39347, 39354, 39362, and 39731. The expiration date range is 2/25/2017-5/1/2017.