Z-0872-2013 Class II Terminated

Recalled by Hill-Rom, Inc. — Batesville, IN

FDA device recall Z-0872-2013 was initiated by Hill-Rom, Inc. on Oct 31, 2012 and is designated Class II. Reason for recall: Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity… The recall status is terminated (terminated Mar 9, 2016). Affected quantity: 3816 in total.

Recall Details

Product Type
Devices
Report Date
Feb 27, 2013
Initiation Date
Oct 31, 2012
Termination Date
Mar 9, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3816 in total

Product Description

Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko ceiling lifts.

Reason for Recall

Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con

Distribution Pattern

Worldwide Distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and CANADA LOCATIONS: AB, BC, MB, NB, NS, ON, and SK.

Code Information

The Liko Overhead Rail System components are not serialized and have been distributed since the early 1980's to the present.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0872-2013
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.