Z-0783-2022 Class II Terminated
Recalled by Hill-Rom, Inc.
— Batesville, IN
FDA device recall Z-0783-2022 was initiated by Hill-Rom, Inc. on Jan 14, 2022 and is designated Class II. Reason for recall: Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately… The recall status is terminated (terminated Nov 7, 2024). Affected quantity: Traverse Rail Carrier E-System, wide 31017XX 5 Traverse Rail Carrier 31025XX 1798.
Recall details
- Product Type
- Devices
- Report Date
- Mar 23, 2022
- Initiation Date
- Jan 14, 2022
- Termination Date
- Nov 7, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- Traverse Rail Carrier E-System, wide 31017XX 5 Traverse Rail Carrier 31025XX 1798
Product description
(1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX component to Liko overhead lift. Intended for to make safe lifting of a patient possible. Affected Devices: 3101704 TRAV RAIL CARR E-SYST STD 3102514 TR CARRIER SLIM/RAISED, LR 3101705 TRAV RAIL CARR ESYST WIDE 3102517 TR CARR. NORMAL/LOWER. LR 3102506 TR CARRIER WIDE/RAISED ML 3102519 TR CARR. LOW.LR 30MM/PAIR 3102511 TR CARRIER NORMAL, LR 3102531 TRAVERSE RL CARRIER STD/RAISED 3102512 TR CARRIER WIDE, LR 3102532 TRAVERSE RL CARRIER WD/RAISED 3102513 TR CARRIER SLIM, LR 3102562 TR CARRIER, PARALLEL/WIDE
Reason for recall
Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients
Distribution pattern
Worldwide distribution - US Nationwide and the countries of AT, CA, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IT, JP, NL, NO, NZ, RU, SE, SG, SK.
Code information
Distributed between 04JAN2021 and 24MAR2021. P/N Product name 3130001 Multirall 200 3301030 LikoGuard L 3301040 LikoGuard XL 3121001 Likorall 200 3122005 Likorall 242 ES natural 3122005NA Likorall 242 ES natural 3122006 Likorall 242 ES white 3122006NA Likorall 242 ES white 3122009 Likorall 242 S natural 3122009NA Likorall 242 S natural 3122010 Likorall 242 S white 3122010NA Likorall 242 S white 3122007 Likorall 242 ES R2R, natural 3122007NA Likorall 242 ES R2R, natural 3122008 Likorall 242 ES R2R, white 3122008NA Likorall 242 ES R2R, white 3122011 Likorall 242 S R2R, natural 3122011NA Likorall 242 S R2R, natural 3122012 Likorall 242 S R2R, white 3122012NA Likorall 242 S R2R, white 3123001 Likorall 243 ES, natural 3123001NA Likorall 243 ES, natural 3123002 Likorall 243 ES, white 3123002NA Likorall 243 ES, white 3122501 Likorall 250 ES, natural 3122501NA Likorall 250 ES, natural 3122502 Likorall 250 ES, white 3122502NA Likorall 250 ES, white 3124050 Likorall 250 S, IRC, natural
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0783-2022
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Hill-Rom, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key