Z-0878-2023 Class II Ongoing
FDA device recall Z-0878-2023 was initiated by Ortho-Clinical Diagnostics, Inc. on Nov 17, 2022 and is designated Class II. Reason for recall: Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results The recall status is ongoing. Affected quantity: 4,158 units.
Recall Details
- Product Type
- Devices
- Report Date
- Jan 11, 2023
- Initiation Date
- Nov 17, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,158 units
Product Description
VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the quantitative measurement of intact parathyroid hormone (iPTH) in human serum and plasma (EDTA or heparin) Catalog Number: 6802892
Reason for Recall
Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Bermuda HM12, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Code Information
UDI-DI: 6802892 (10758750006267) Lot Numbers/Exp Date: 1610 24-Apr-2023; 1621 24-Apr-2023; 1630 08-May-2023; 1640 08-May-2023; 1645 08-May-2023; 1650 29-May-2023; 1670 06-Jun-2023
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0878-2023
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab