Z-0878-2023 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, Inc. — Rochester, NY

FDA device recall Z-0878-2023 was initiated by Ortho-Clinical Diagnostics, Inc. on Nov 17, 2022 and is designated Class II. Reason for recall: Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results The recall status is ongoing. Affected quantity: 4,158 units.

Recall Details

Product Type
Devices
Report Date
Jan 11, 2023
Initiation Date
Nov 17, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,158 units

Product Description

VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the quantitative measurement of intact parathyroid hormone (iPTH) in human serum and plasma (EDTA or heparin) Catalog Number: 6802892

Reason for Recall

Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Bermuda HM12, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Code Information

UDI-DI: 6802892 (10758750006267) Lot Numbers/Exp Date: 1610 24-Apr-2023; 1621 24-Apr-2023; 1630 08-May-2023; 1640 08-May-2023; 1645 08-May-2023; 1650 29-May-2023; 1670 06-Jun-2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0878-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.