Recalled by Hospira Inc. — Lake Forest, IL

FDA device recall Z-1016-2017 was initiated by Hospira Inc. on Feb 19, 2013 and is designated Class II. Reason for recall: Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of… The recall status is terminated (terminated Apr 24, 2017). Affected quantity: 48,050 units.

Recall Details

Product Type
Devices
Report Date
Jan 18, 2017
Initiation Date
Feb 19, 2013
Termination Date
Apr 24, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
48,050 units

Product Description

Minibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For the administration of drugs and solutions to a patient's vascular system through a vascular access device.

Reason for Recall

Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.

Distribution Pattern

Nationwide Distribution

Code Information

Product List Number: 14699-28; Lot Number: 17-077-NS

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1016-2017
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.