Z-1016-2026 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 28, 2026
- Initiation Date
- November 26, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15581 units
Product Description
MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO
Reason for Recall
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Distribution Pattern
US, CA, PA, GUAM, VI
Code Information
UDI/DI 40080196325901, All lots