Z-1021-2013 Class II Terminated

Recalled by Microbiologics Inc — Saint Cloud, MN

FDA device recall Z-1021-2013 was initiated by Microbiologics Inc on Feb 11, 2013 and is designated Class II. Reason for recall: Microbiologics Inc is initiating a voluntary recall on products KWIK-QC Acid Fast Stain Slide, SL42-10 and KWIK-QC Mycobacterium, SL43-10. These products are being recalled because the control… The recall status is terminated (terminated Mar 10, 2016). Affected quantity: 396 boxes of 10 slides each.

Recall Details

Product Type
Devices
Report Date
Apr 3, 2013
Initiation Date
Feb 11, 2013
Termination Date
Mar 10, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
396 boxes of 10 slides each.

Product Description

Microbiologics KWIK-QC SLIDES ( KWIK-QC Acid Fast Stain Slides and KWIK-QC Mycobacterium Slides). Slide preparations serving as quality control challenges to demonstrate satisfactory performance of staining reagents.

Reason for Recall

Microbiologics Inc is initiating a voluntary recall on products KWIK-QC Acid Fast Stain Slide, SL42-10 and KWIK-QC Mycobacterium, SL43-10. These products are being recalled because the control organisms are not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain. This contamination could pote

Distribution Pattern

Worldwide Distribution-USA including the states of CA, IN, MS, NY, OH, PA, and SD, and the countries of AUSTRALIA, AUSTRIA, UNITED ARAB EMIRATES, MALAYSIA, TAIWAN, JAPAN, TURKEY, BELGIUM, COLOMBIA, CANADA, SOUTH AFRICA, and VENEZUELA.

Code Information

SL42-10 Lots 4239 to 4243 SL43-10 Lots 4333 to 4337

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1021-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.