Z-1029-2017 Class II Terminated

Recalled by Fresenius Medical Care Renal Therapies Group, LLC — Waltham, MA

Recall Details

Product Type
Devices
Report Date
January 25, 2017
Initiation Date
December 21, 2016
Termination Date
March 17, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1400 units (US) 556 units (OUS)

Product Description

Fresenius 2008K@home Series : Hemodialysis System

Reason for Recall

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

Distribution Pattern

Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam

Code Information

All serial numbers